Hi-Torque Versacore™

Peripheral Workhorse Guide Wires

From Start To Finish

Performance

  • Workhorse wire allows for delivery of PTA catheters and stents
  • Core-to-tip design enables 1:1 torque1
  • Combination of support and flexibility meets most peripheral intervention needs
  • Polyethylene jacket with silicon coating
  • Soft and shapeable tip has gold design for enhanced visibility
  • Guide wire comes in lengths of 145 cm, 175 cm, 260 cm and 300 cm

Three Different Tip Types

  • Hi-Torque Versacore™ Modified J Guide Wire configuration is a highly versatile guide wire, which combines support and flexibility for successful access to a wide range of vascular anatomy.
  • Hi-Torque Versacore™ Floppy Guide Wire configuration is a highly flexible guide wire for atraumatic negotiation of high-grade stenoses and multiple lesions.
  • Hi-Torque Versacore™ Standard Guide Wire configuration is a highly supportive guide wire that offers increased catheter support and trackability for traversing acute angles.

Ordering Information

ProductPart NumberDiameterLengthTip StyleUnits Per Package
Hi-Torque Versacore™ Modified J1012068-010.035 in145 cmCore-To-Tip5
Hi-Torque Versacore™ Modified J1012068-050.035 in175 cmCore-To-Tip5
Hi-Torque Versacore™ Modified J1012068-090.035 in260 cmCore-To-Tip5
Hi-Torque Versacore™ Modified J1012068-070.035 in300 cmCore-To-Tip5
Hi-Torque Versacore™ Floppy1012068-020.035 in145 cmCore-To-Tip5
Hi-Torque Versacore™ Floppy1012068-040.035 in175 cmCore-To-Tip5
Hi-Torque Versacore™ Floppy1012068-060.035 in260 cmCore-To-Tip5
Hi-Torque Versacore™ Standard1012068-030.035 in145 cmCore-To-Tip5
Hi-Torque Versacore™ Standard1012068-080.035 in300 cmCore-To-Tip5
LOC™ Guide Wire Extension1012068-100.035 in115 cmN/A5

Full Peripheral Guide Wire Brochure

Customer Service

Data on file at Abbott.

References

  1. Tóth et al. How to select a guidewire: technical features and key characteristics. Heart 2015;101:645-652

MAT-2635317 v3.0

Important Safety Information

Hi-Torque™ Versacore
Guide Wire

rx-only-logo

Indications

The Hi-Torque Versacore™ Guide Wire is indicated for use in angiographic procedures to introduce and position diagnostic and interventional devices within the peripheral vasculature during percutaneous procedures. The wire can be torqued to facilitate navigation through tortuous vessels.
 

Contraindications

The Hi-Torque Versacore™ Guide Wire is not intended for use in the coronary or cerebral vasculature.
 

Warnings

This device is not designed for use with atherectomy devices.

The lifetime of the device is the labeled shelf life. This device is designed and intended for one time use only. Do not resterilize and / or reuse. The safety and effectiveness of this device have not been established after being reprocessed for multiple uses.

Use extreme caution and careful judgement in patients for whom anticoagulation is not indicated.

If contrast agents are used, use extreme caution in patients who have had a severe reaction to contrast agents and who cannot be adequately premedicated.

Persons with known history of allergies to any of the components of this device listed below may suffer an allergic reaction to this guide wire: stainless steel, polytetrafluoroethylene (PTFE) coating, gold-silver-copper-zinc alloy, silver-copper-zinc-nickel alloy, high density polyethylene, low density polyethylene, silicone coating, and cyanoacrylate ester. Prior to its use on the patient, the patient should be counseled on the materials contained in the device, and a thorough history of allergies must be discussed.

This device has a hydrophobic coating at the proximal end and a hydrophobic coating on the distal end of the device that enhances the penetration capacity and provides good torque control for lengths as specified in the table below:

 

Product

Type of Coating and Coating Length at the Proximal End

Type of Coating and Coating Length at Both Proximal and Distal Ends

Hi-Torque Versacore, 145 cm

Silicone coating (Hydrophobic coating), 45.0 cm

 

 

PTFE coating (Hydrophobic coating) over coil, 91.0 cm starting from the distal end

Hi-Torque Versacore, 175 cm

Silicone coating (Hydrophobic coating), 75.0 cm

Hi-Torque Versacore, 260 cm

Silicone coating (Hydrophobic coating), 160.0 cm

Hi-Torque Versacore, 300 cm

Silicone coating (Hydrophobic coating), 200.0 cm


Refer to section Preparation for Use for further information on how to prepare and use this device to ensure it performs as intended. Failure to abide by the warnings in this labeling might result in damage to the device coating, which may necessitate intervention or result in serious adverse events.

Carefully observe the instructions under “Do Not” and “Do” below. Failure to do so may result in vessel trauma, guide wire damage, guide wire tip separation, or stent damage. If resistance is observed at any time, determine the cause under fluoroscopy and take remedial action as needed. Use the most suitable guide wire for the lesion being treated.

Do Not:

  • Push, auger, withdraw, or torque a guide wire that meets resistance.
  • Torque a guide wire if the tip becomes entrapped within the vasculature.
  • Allow the guide wire tip to remain in a prolapsed condition.
  • Deploy a stent such that it will entrap the wire between the vessel wall and the stent.

Do:

  • Advance or withdraw the guide wire slowly.
  • Use the radiopaque marker of the interventional device to confirm position.
  • Examine the tip movement under fluoroscopy before manipulating, moving, or torquing the guide wire.
  • Observe the wire under fluoroscopy for tip buckling, which is a sign of resistance.
  • Maintain continuous flush while removing and reinserting the guide wire to prevent air from entering the catheter system. Perform exchanges slowly to prevent air entry and / or trauma.
  • When reintroducing the guide wire, confirm that the interventional device tip is free within the vessel lumen and that the tip is parallel to the vessel wall.
  • Use extreme caution when moving a guide wire through a non-endothelialized stent, or through stent struts, into a bifurcated vessel. Use of this technique involves additional patient risks, including the risk that the wire may become caught on the stent strut.
  • Consider that if a secondary wire is placed in a bifurcation branch, this wire may need to be retracted prior to stent deployment because there is additional risk that the secondary wire may become entrapped between the vessel wall and the stent.


Limitations

PTA should only be performed at centers where emergency peripheral artery bypass surgery is available.
 

Precautions

Never attach the torque device to the modified portion of the proximal end of the extendable guide wire; otherwise, guide wire damage may occur, preventing the ability to attach the Abbott-compatible Guide Wire Extension.

Guide wires are delicate instruments and should be handled carefully. Prior to use and when possible during the procedure, inspect the guide wire carefully for bends, kinks, or other damage. Do not use damaged wires. Using a damaged wire may result in vessel damage and / or inaccurate torque response.

Confirm the compatibility of the guide wire diameter with the interventional device before actual use.

Refer to the indication on the label and the Instructions for Use to confirm the appropriate vasculature that this guide wire may be used in. Failure to abide by the above recommendation may result in size mismatch of blood vessel and guide wire, which can result in vessel injury, such as, but not limited to, perforation, dissection, rupture, and avulsion.

Free movement of the guide wire within the interventional device is an important feature of a steerable guide wire system because it gives the user valuable tactile information. Test the system for any resistance prior to use. Adjust or replace the hemostatic valve with an adjustable valve if it is found to inhibit guide wire movement.

It is recommended that the user determine the source of resistance, exercise caution when removing the device and / or other components as a unit, and exchange the device for a new one to complete the procedure.

Avoid abrasion of the coating. Do not withdraw or manipulate the coated guide wire through a metal cannula or sharp-edged object. Manipulation, advancement, and / or withdrawal through a metal device may result in destruction and / or separation of the outer coating, which may cause coating material to remain in the vasculature. This in turn may lead to unintended adverse events requiring additional intervention.

After use, this product may be a potential biohazard. Handle and dispose of it in accordance with accepted medical practice and applicable local, state, and federal laws and regulations.

The integrity and performance of the device coating can be negatively impacted by preparation with incompatible media or solvents. Please take note of the following important recommendations:

  • Avoid wiping the device with dry gauze as this may damage the device coating.
  • Avoid excessive wiping of the coated devices.
  • Avoid using alcohol, antiseptic solutions, or other solvents to pre-treat the device because this may cause unpredictable changes in the coating which could negatively affect the safety and performance of the guide wire.

One or more components of this device may contain the following substance defined as

CMR 1B in a concentration above 0.1% weight by weight: Cobalt; Chemical Abstracts Service (CAS) No. 7440-48-4; EC No. 231-158-0. Current scientific evidence supports that medical devices manufactured from stainless steel alloys containing cobalt do not cause an increased risk of cancer or adverse reproductive effects.

The safety and effectiveness of this Hi-Torque Versacore™ Guide Wire have not been established in the pediatric population.
 

Adverse Events

Potential adverse events associated with use of this device may include the following but not limited to:

  • Allergic reaction or hypersensitivity to latex, contrast agent, anesthesia, device materials, and drug reactions to anticoagulation, or antiplatelet drugs
  • Vascular access complications which may require transfusion or vessel repair, including:
    • Bleeding (ecchymosis, oozing, hematoma, hemorrhage, retroperitoneal hemorrhage)
    • Arteriovenous fistula, pseudoaneurysm, aneurysm, dissection, perforation / rupture, and laceration
    • Embolism (air, tissue, plaque, thrombotic material, or device)
  • Target artery complications which may require additional intervention, including:
    • Total occlusion or abrupt closure
    • Arteriovenous fistula, pseudoaneurysm, aneurysm, dissection, perforation / rupture
    • Embolism (air, tissue, plaque, thrombotic material, or device)
    • Artery or stent thrombosis
    • Stenosis or restenosis
    • Vessel spasm
    • Claudication
  • Venous thromboembolism (including pulmonary embolism)
  • Hypotension / hypertension
  • Peripheral nerve injury, neuropathy
  • Other ischemic conditions / infarct
    • Tissue / organ ischemia
    • Tissue necrosis
    • Ulcer
    • Acute limb ischemia
  • Infection – local and systemic (including post-procedural)
    • Abscess
    • Sepsis / infection including bacteremia / cellulitis / septicemia
    • Contrast-induced renal insufficiency or renal failure
    • Death