Amplatzer™ Vascular Plug (AVP)
Designed to provide precise positioning in short landing zones1
- Single lobe, single layer device2
- Platinum marker bands on both ends2
- Select a device that is ~30-50% larger in diameter than the target vessel diameter2


Device Description2:
- Self-expanding device made from a Nitinol wire mesh
- The device is shipped attached to a 135cm nitinol delivery wire in a hoop dispenser
- The device is preloaded in a loader
- No plastic vise is provided
- Sizes ranging from 4mm – 16mm in 2mm increments
Materials2:
- Plug: Nitinol
- Delivery wire: Nitinol
- Marker bands: Platinum
- Micro screw: Stainless steel

References
- Lopera, Jorge E. “The Amplatzer vascular plug: review of evolution and current applications.” Seminars in interventional radiology. Vol. 32. No. 04. Thieme Medical Publishers, 2015.
- Amplatzer™ Vascular Plug and Vascular Plug II - Instructions for Use
MAT-2002814 v1.0
Important Safety Information
AMPLATZER™ Vascular Plug

Intended Use: The AMPLATZER™ Vascular Plug is indicated for arterial and venous embolizations in the peripheral vasculature.
Contraindications: None known.
Warnings: The safety and effectiveness of this device for cardiac uses (for example, patent ductus arteriosus or paravalvular leak closures) and neurological uses have not been established.
Precautions:
Handling
- Patients allergic to nickel may suffer an allergic reaction to this device.
- This device was sterilized with ethylene oxide and is for single use only. Do not reuse or resterilize this device. Attempts to resterilize this device can cause a malfunction, insufficient sterilization, or harm to the patient.
- Use on or before the last day of the expiration month that is printed on the product packaging label.
- Do not use this device if the sterile package is open or damaged.
- This device should be used only by physicians who are trained in standard endovascular techniques. The physician should determine which patients are candidates for procedures that use this device.
- Store in a dry place.
MRI Safety Information:
![]() | |
Device Name | Amplatzer™ Vascular Plug |
Static Magnetic Field Strength (B0) | 1.5 T or 3.0 T |
Maximum Field Spatial Gradient | 19T/m (1900 gauss/cm) |
RF Excitation | Circularly Polarized (CP) |
RF Transmit Coil Type | Body Coil |
Operating Mode | Normal Operating Mode |
Maximum Whole Body SAR | 2.0W/kg (Normal Operating Mode) |
Maximum Head SAR | N/A |
Scan Duration | 2.0W/kg whole-body-averaged SAR for 15 minutes of continuous scanning |
MR Image Artifact | The presence of this implant may produce an image artifact. |
Potential Adverse Events: Potential complications include, but are not limited to: death, embolization of the device, hematoma at the site of entry, stroke, or vessel perforation, or embolization of the device.