AMPLATZER™ VASCULAR PLUG II
(AVP II)

Sizing and Specs1

Rapid embolization2 and can adjust to variable landing zones3

A - Vascular Plug Diameter

B - Unconstrained Length

Sizing and Device Selection

Amplatzer™ Vascular Plug II

 

Vessel Size Device Dimensions Delivery Catheter Requirements*
Treatable Vessel Diameter Range Model/Reorder Number Vascular Plug Diameter (A) Unconstrained Length (B) Minimum and Maximum Internal Diameter Minimum Sheath Size or Minimum Guide Catheter Size Maximum Delivery Catheter Length**
2.0 mm-2.5 mm 9-AVP2-003 3mm 6mm 1.42-1.70 mm / 0.056-0.067 in ≥4 F   ≥5 F ≤ 100 cm
2.5 mm-3.0 mm 9-AVP2-004 4mm 6mm 1.42-2.50 mm / 0.056-0.098 in ≥4 F   ≥5 F ≤ 100 cm
4.0 mm-4.5 mm 9-AVP2-006 6mm 6mm 1.42-2.50 mm / 0.056-0.098 in ≥4 F   ≥5 F ≤ 100 cm
5.5 mm-6.0 mm 9-AVP2-008 8mm 7mm 1.42-2.69 mm / 0.056-0.106 in ≥4 F   ≥5 F ≤ 100 cm
6.5 mm-7.5 mm 9-AVP2-010 10mm 7mm 1.78-2.69 mm / 0.070-0.106 in ≥5 F   ≥6 F ≤ 100 cm
8.0 mm-9.0 mm 9-AVP2-012 12mm 9mm 1.78-2.69 mm / 0.070-0.106 in ≥5 F   ≥6 F ≤ 100 cm
9.5 mm-11.0 mm 9-AVP2-014 14mm 10mm 2.18-2.69 mm / 0.086-0.106 in ≥6 F   ≥8 F ≤ 100 cm
10.5 mm-12.5 mm 9-AVP2-016 16mm 12mm 2.18-2.69 mm / 0.086-0.106 in ≥6 F   ≥8 F ≤ 100 cm
12.0 mm-14.0 mm 9-AVP2-018 18mm 14mm 2.49-2.69 mm / 0.098-0.106 in ≥7 F   ≥9 F ≤ 100 cm
13.5 mm-15.5 mm 9-AVP2-020 20mm 16mm 2.49-2.69 mm / 0.098-0.106 in ≥7 F   ≥9 F ≤ 100 cm
14.5 mm-17.0 mm 9-AVP2-022 22mm 18mm 2.49-2.69 mm / 0.098-0.106 in ≥7 F   ≥9 F ≤ 100 cm

Treatable vessel diameter range based on the devices Instructions for Use to select a plug that is oversized by approximately 30-50% at the occlusion site.
*The Amplatzer™ Vascular Plug II is delivered utilizing either a sheath or guide catheter meeting the minimum internal diameter requirements above.
**Each Amplatzer™ Vascular Plug II comes pre-loaded on a 135cm nitinol delivery wire.

Amplatzer™ Family of Vascular Plugs

  • Amplatzer™ Vascular Plug (AVP)
  • Amplatzer™ Vascular Plug II (AVP II)
  • Amplatzer™ Vascular Plug 4 (AVP 4)

References

  1. Amplatzer™ Vascular Plug and Vascular Plug II - Instructions for Use
  2. Pech M, Kraetsch A, Winers G, et al., Embolization of the Gastroduodenal Artery Before Selective Internal Radiotherapy: A Prospectively Randomized Trial Comparing Platinum-Fibered Microcoils with the Amplatzer Vascular Plug II. CVIR 2009(32)3:455-61.
  3. AVP II is available in 8 unconstrained lengths ranging from 6 -18 mm.

MAT-2002813 v1.0

DO YOU WISH TO CONTINUE AND EXIT CARDIOVASCULAR.ABBOTT?

CONTENTS OF THE SITE ARE NOT UNDER THE CONTROL OF ABBOTT.

True
accessibility
© 2016 Abbott. All Rights Reserved. Please read the Legal Notice for further details.

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

accessibility

DO YOU WISH TO CONTINUE AND EXIT CARDIOVASCULAR.ABBOTT?

CONTENTS OF THE SITE ARE NOT UNDER THE CONTROL OF ABBOTT.