AMPLATZER™ VASCULAR PLUG 4
(AVP 4)

Sizing and Specs1

Low-profile embolization compatible with 0.038" guide wire compatible diagnostic catheters1 that reach distal vasculature through tortuous anatomy.2

AVP 4 sizing

A - Vascular Plug Diameter

B - Unconstrained Length

Sizing and Device Selection

Amplatzer™ Vascular Plug 4

 

Vessel Size Device Dimensions Diagnostic Catheter Requirements*
Treatable Vessel Diameter Range Model/Reorder Number Vascular Plug Diameter (A) Unconstrained Length (B) Requirements for diagnostic catheters** Maximum Diagnostic Catheter Length***
2.5 mm-3.0 mm 9-AVP038-004 4 mm 10.0 mm 0.038" Guidewire-Compatible Diagnostic Catheter ≤ 125 cm
3.5 mm-4.0 mm 9-AVP038-005 5 mm 10.5 mm 0.038" Guidewire-Compatible Diagnostic Catheter ≤ 125 cm
4.0 mm-4.5 mm 9-AVP038-006 6 mm 11.0 mm 0.038" Guidewire-Compatible Diagnostic Catheter ≤ 125 cm
4.5 mm-5.5 mm 9-AVP038-007 7 mm 12.5 mm 0.038" Guidewire-Compatible Diagnostic Catheter ≤ 125 cm
5.5 mm-6.0 mm 9-AVP038-008 8 mm 13.5 mm 0.038" Guidewire-Compatible Diagnostic Catheter ≤ 125 cm

Treatable vessel diameter range based on the devices Instructions for Use to select a plug that is oversized by approximately 30-50% at the occlusion site.
*The Amplatzer™ Vascular Plug 4 is delivered utilizing a Diagnostic Catheter meeting the minimum internal diameter requirements above.
**The Amplatzer™ Vascular Plug 4 has been tested for compatibility with the following diagnostic catheters and corresponding lengths:
    5F Dx Catheters: Boston Scientific IMAGER‡ II (100 cm or less) and Merit Medical Impress‡ (125 cm or less)
    4F Dx Catheters: Cordis‡ TEMPO‡ or TEMPO‡ AQUA‡ (100 cm or less)
***Each Amplatzer™ Vascular Plug 4 comes pre-loaded and packaged on an 0.038" x 155cm-long PTFE-coated delivery wire.

Amplatzer™ Family of Vascular Plugs

  • Amplatzer™ Vascular Plug (AVP)
  • Amplatzer™ Vascular Plug II (AVP II)
  • Amplatzer™ Vascular Plug 4 (AVP 4)

References

  1. Amplatzer™ Vascular Plug 4 – Instructions for Use
  2. Lopera, Jorge E. “The Amplatzer vascular plug: review of evolution and current applications.” Seminars in interventional radiology. Vol. 32. No. 04. Thieme Medical Publishers, 2015.

MAT-2002813 v1.0

DO YOU WISH TO CONTINUE AND EXIT CARDIOVASCULAR.ABBOTT?

CONTENTS OF THE SITE ARE NOT UNDER THE CONTROL OF ABBOTT.

True
accessibility
© 2016 Abbott. All Rights Reserved. Please read the Legal Notice for further details.

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

accessibility

DO YOU WISH TO CONTINUE AND EXIT CARDIOVASCULAR.ABBOTT?

CONTENTS OF THE SITE ARE NOT UNDER THE CONTROL OF ABBOTT.