Abbott is notifying customers of a voluntary recall of the AC Power Cord (model 107760) for the HeartMate™ Mobile Power Unit (MPU; model 107754). Some MPU AC power cords did not fully lock into place when inserted into the MPU and could be pulled from the MPU without pressing the yellow button on the plug. This issue has been linked to nonconforming power cords packaged with certain MPUs distributed in May 2025 and June 2025. Abbott has asked that clinicians review the issue with impacted patients and provide a patient letter as applicable and discontinue the use of impacted power cords, but not recommending removal of the MPU from the field at this time.
Check for a Device
Check to see if a specific mobile power unit AC Power Cord is part of the AC Power Cords for HeartMate Mobile Power Unit Device Recall.


Pictures are for reference only to locate the Serial, Reference or Lot number
Device Lookup Instructions
Please follow the steps below to determine if a device is subject to this recall.
- Enter one of the numbers above. No devices beginning in "MPU-" are in scope of this Recall.
- If the device is subject to this recall, review the recommendations in the customer or patient letter.
- If the device is not subject to this recall, no further action is needed.