MRI Safety

The Gallant™ ICD is MR Conditional when used with MR Conditional leads for full body scans using a 1.5T or 3T field strength MRI scanner.

For additional information about specific MR Conditional devices and leads, including scan parameters, warnings, precautions, adverse conditions to MRI scanning and potential adverse events, please refer to the Abbott MRI Ready Systems Manual.

MRI Scan Parameters

Lead Model Magnet (Tesla) RF Transmit Conditions Scan Region
Durata™ Defibrillation Lead
7120Q (lead lengths: 58, 65 cm)
7122Q (lead lengths: 58, 65 cm)
1.5T / 3T Normal Operating Mode Full-body
Optisure™ Defibrillation Lead
LDA220Q (lead lengths: 58, 65 cm)
LDA210Q (lead lengths: 58, 65 cm)
1.5T / 3T
Tendril™ STS Pacing Lead
2088TC (lead lengths: 46, 52 cm)
1.5T / 3T
Tendril MRI™ Pacing Lead
LPA1200M (lead lengths: 46, 52 cm)
1.5T
MRI Settings*
TACHY Therapy Disabled
MRI Mode DOO; VOO; AOO; Pacing Off
MRI Base Rate 30-100 bpm
MRI Paced AV Delay 25-120 ms
MRI Pulse Amplitude 5.0 or 7.5 V
MRI Pulse Width 1.0 ms
MRI Pulse Configuration Bipolar
MRI Timeout Off; 3; 6; 9; 12; 24 hours

*DOO, AOO and paced AV delays are only applicable to the DR models.

This device is commercially available for use in select international markets.

MAT-2008639 v1.0

DO YOU WISH TO CONTINUE AND EXIT CARDIOVASCULAR.ABBOTT?

CONTENTS OF THE SITE ARE NOT UNDER THE CONTROL OF ABBOTT.

False
accessibility
© 2021 Abbott. All Rights Reserved. Please read the Legal Notice for further details.

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

accessibility

DO YOU WISH TO CONTINUE AND EXIT CARDIOVASCULAR.ABBOTT?

CONTENTS OF THE SITE ARE NOT UNDER THE CONTROL OF ABBOTT.