1
CARDIOVASCULAR
hamburger
AVP II

AMPLATZER™ VASCULAR PLUG II
(AVP II)

Rapid embolization¹ and can adjust to variable landing zones

  • Multi-layered mesh designed to increase density and flow disturbance
  • Six planes of cross sectional coverage resulting in rapid occlusion1
  • Select a device that is ~30-50% larger in diameter than the target vessel diameter
amplatzer vascular plug II
amplatzer vascular plug II

Device Description:

  • Self-expanding device made from a Nitinol wire mesh
  • The device is shipped attached to a 135cm nitinol delivery wire in a hoop dispenser
  • The device is preloaded in a loader
  • A plastic vise is also included and may be attached to the delivery wire to facilitate device detachment

Materials:

  • Plug: Nitinol
  • Delivery wire: Nitinol
  • Marker bands: Platinum
  • Micro screw: Stainless steel
amplatzer vascular plug ii materials

Sizing and Specs

Rapid embolization¹ and can adjust to variable landing zones

amplatzer vascular plug II

Sizing and Device Selection

AMPLATZER Vascular Plug II

 

Vessel Size Device Dimensions Delivery Catheter Requirementsc
Treatable Vessel Diameter Range (mm) Model/Reorder Number Vascular Plug Diameter (A) Unconstrained Length (B) Minimum and Maximum Internal Diameter (mm/in) Minimum Sheath Size (F) or Minimum Guide Catheter Size (F) Maximum Delivery Catheter Lengthd (cm)
2.0 mm-2.5 mm 9-AVP2-003 3mm 6mm 1.42-1.70 mm / 0.056-0.067 in ≥4 F   ≥5 F ≤ 100 cm
2.5 mm-3.0 mm 9-AVP2-004 4mm 6mm 1.42-2.49 mm / 0.056-0.098 in ≥4 F   ≥5 F ≤ 100 cm
4.0 mm-4.5 mm 9-AVP2-006 6mm 6mm 1.42-2.49 mm / 0.056-0.098 in ≥4 F   ≥5 F ≤ 100 cm
5.5 mm-6.0 mm 9-AVP2-008 8mm 7mm 1.42-2.69 mm / 0.056-0.106 in ≥4 F   ≥5 F ≤ 100 cm
6.5 mm-7.5 mm 9-AVP2-010 10mm 7mm 1.78-2.69 mm / 0.070-0.106 in ≥5 F   ≥6 F ≤ 100 cm
8.0 mm-9.0 mm 9-AVP2-012 12mm 9mm 1.78-2.69 mm / 0.070-0.106 in ≥5 F   ≥6 F ≤ 100 cm
9.5 mm-11.0 mm 9-AVP2-014 14mm 10mm 2.18-2.69 mm / 0.086-0.106 in ≥6 F   ≥8 F ≤ 100 cm
10.5 mm-12.5 mm 9-AVP2-016 16mm 12mm 2.18-2.69 mm / 0.086-0.106 in ≥6 F   ≥8 F ≤ 100 cm
12.0 mm-14.0 mm 9-AVP2-018 18mm 14mm 2.49-2.69 mm / 0.098-0.106 in ≥7 F   ≥9 F ≤ 100 cm
13.5 mm-15.5 mm 9-AVP2-020 20mm 16mm 2.49-2.69 mm / 0.098-0.106 in ≥7 F   ≥9 F ≤ 100 cm
14.5 mm-17.0 mm 9-AVP2-022 22mm 18mm 2.49-2.69 mm / 0.098-0.106 in ≥7 F   ≥9 F ≤ 100 cm

Treatable vessel diameter range based on the device's Instructions for Use to select a plug that is oversized by approximately 30-50% at the occlusion site.
cThe AMPLATZER Vascular Plug II is delivered utilizing either a sheath or guide catheter meeting the minimum internal diameter requirements above.
dEach AMPLATZER Vascular Plug II comes pre-loaded on a 135 cm nitinol delivery wire.

Deployment

Note: This section should not be used in place of the Instructions for Use. This section only provides an overview of the procedure. Refer to the Instructions for Use for additional information.

Access the vessel and perform an angiogram using standard technique to measure the diameter of the vessel at the desired occlusion site.

Select a vascular plug with a diameter approximately 30-50% larger than the vessel diameter at the occlusion site. Ensure that the occlusion site has sufficient length so the deployed device length will not obstruct other vessels or anatomical structures.

Deployment Steps

amplatzer vascular plug steps
1 Flush hoop dispenser and loader with sterile saline until fluid exits the distal tip to purge air from the loader.
2 Remove device (in loader) and delivery wire from the hoop dispenser.
3 Select a delivery catheter with an internal diameter that will accommodate the selected device. The original access catheter may be used if the internal diameter is sufficient for the selected device. Prepare the catheter for use (according to the manufacturer’s instructions for use).
4 Advance the delivery catheter over the guidewire until the distal tip of the catheter is at the distal edge of the occlusion site (according to the manufacturer’s instructions for use).
5 Remove the guidewire.
6 Insert loader into the delivery catheter through a Y-connector or hemostasis valve.
7 Allow blood backflow or aspirate the system to ensure air is purged from the catheter and loader.
8 Do not over tighten Y-connector to avoid damaging loader.
9 Advance the delivery wire to advance the device to the distal end of the delivery catheter. Do not twist or rotate the delivery wire during advancement to ensure device does not prematurely detach.
10 Hold delivery wire in place and slowly retract the delivery catheter to deploy the device at the occlusion site.
11 "Verify position of the device using the radiopaque marker bands.
~ If unsatisfactory, stabilize the wire and re-advance the delivery catheter until the device is completely within the catheter. Reposition and deploy, or remove the device from the patient.
~ If satisfactory, attach the plastic visea to the wire and rotate the delivery wire counterclockwise until it separates from the device."
12 Remove delivery catheter and wire from the patient.
Steps 1-3
amplatzer vascular plug steps
amplatzer vascular plug steps
amplatzer vascular plug steps
Steps 4-5
amplatzer vascular plug steps
amplatzer vascular plug steps
Steps 6-9
amplatzer vascular plug steps
amplatzer vascular plug steps
Steps 10-12
amplatzer vascular plug steps
amplatzer vascular plug steps
amplatzer vascular plug steps

Ordering

Sizing and Device Selection

AMPLATZER Vascular Plug II

amplatzer vascular plug II sizing
Device Dimensions Vessel Size
Model/ Reorder Number Vascular Plug Diameter[A] Unconstrained Length [B] Product Descriptionb Treatable Vessel Diameter Range (mm)
9-AVP2-003 3 mm 6 mm AMPLATZER™ Vascular Plug II - 3mm x 6mm 2.0 mm-2.5 mm
9-AVP2-004 4 mm 6 mm AMPLATZER™ Vascular Plug II - 4mm x 6mm 2.5 mm-3.0 mm
9-AVP2-006 6 mm 6 mm AMPLATZER™ Vascular Plug II - 6mm x 6mm 4.0 mm-4.5 mm
9-AVP2-008 8 mm 7 mm AMPLATZER™ Vascular Plug II - 8mm x 7mm 5.5 mm-6.0 mm
9-AVP2-010 10 mm 7 mm AMPLATZER™ Vascular Plug II - 10mm x 7mm 6.5 mm-7.5 mm
9-AVP2-012 12 mm 9 mm AMPLATZER™ Vascular Plug II - 12mm x 9mm 8.0 mm-9.0 mm
>9-AVP2-014 14 mm 10 mm AMPLATZER™ Vascular Plug II - 14mm x 10mm 9.5 mm-11.0 mm
9-AVP2-016 16 mm 12 mm AMPLATZER™ Vascular Plug II - 16mm x 12mm 10.5 mm-12.5 mm
9-AVP2-018 18 mm 14 mm AMPLATZER™ Vascular Plug II - 18mm x 14mm 12.0 mm-14.0 mm
9-AVP2-020 20 mm 16 mm AMPLATZER™ Vascular Plug II - 20mm x 16mm 13.5 mm-15.5 mm
9-AVP2-022 22 mm 18 mm AMPLATZER™ Vascular Plug II - 22mm x 18mm 14.5 mm-17.0 mm

Amplatzer Family of Vascular Plugs

MAT-2002814 v1.0

DO YOU WISH TO CONTINUE AND EXIT CARDIOVASCULAR.ABBOTT?

CONTENTS OF THE SITE ARE NOT UNDER THE CONTROL OF ABBOTT.

False
accessibility
© 2019 Abbott. All Rights Reserved. Please read the Legal Notice for further details.

Unless otherwise specified, all product and service names appearing in this Internet site are trademarks owned by or licensed to Abbott, its subsidiaries or affiliates. No use of any Abbott trademark, trade name, or trade dress in this site may be made without the prior written authorization of Abbott, except to identify the product or services of the company.

accessibility

DO YOU WISH TO CONTINUE AND EXIT CARDIOVASCULAR.ABBOTT?

CONTENTS OF THE SITE ARE NOT UNDER THE CONTROL OF ABBOTT.